Skip to content

Common Issues

Compatible with: LRS4000 multiScan100 multiScan200 picoScan100

Quick fixes for the problems you are most likely to encounter during commissioning and integration. Each row links to the page with the details. Available settings, protocols, and menu paths vary by device family and firmware. Use the relevant device guide and release notes to confirm support.

Connection

Connection: symptom and check.

Symptom Check
Sensor not reachable / ping fails Computer and sensor in the same subnet; power and Ethernet connected (Getting Started)
IP address unknown Device Discovery: SOPAS ET search, or ARP when device search is disabled
Lost connection after an IP change Move the computer to the new subnet, or use device search to find the sensor
IP change through device search is rejected The default IP change policy allows it only within 60 s after boot; reboot or log in (IP Address Assignment)
Web UI does not load Open http://<ip>/ in Chrome, Edge, or Firefox; clear the cache; check that the web server is not disabled (Cybersecurity)
HTTPS refuses connections after a certificate upload Read httpsStatus over HTTP on port 80, fix the certificate, and upload it again (HTTPS & Certificates)

Configuration

Configuration: symptom and check.

Symptom Check
Cannot change settings in the web UI Log in with an enabled user level; without login, the UI is read-only
REST response has header.status 4 (Access Denied) The HTTP status is always 200; check header.status. Fetch a new challenge for every POST, check the password and that the user level is enabled, and wait 30 s after five failed logins (REST Interface)
REST file transfer returns 403 Send a valid X-Session-Token header (Backup & Restore)
Settings lost after power cycle Persist with WriteEeprom or Save permanently (REST Interface)
CoLa sFA error response Look up the CoLa error code
Web UI shows an error ID Look up the ID in the Error Codes list
Need to restore factory settings How to perform a factory reset
Firmware update stuck or fails Do not power-cycle while UpdateState is 1; allow 2-5 minutes for flash and reboot; on 3 (Error), check for the correct .spk.signed package (Firmware Update)

Measurement Data

Measurement Data: symptom and check.

Symptom Check
No UDP data arrives Data output enabled and destination IP/port set to your computer; Windows network profile set to Private; firewall allows the UDP port (Compact Format)
No TCP data on LRS4000 or multiScan200 The sensor sends Compact over TCP only while a client is connected; connect to port 2115 (Compact Format)
LMDscandata returns only zeros The output format is not set to LMDscandata; sRN LMDscandata then answers without an error (LMDscandata Format)
Scan view empty Measurement is active and data output is enabled in the device's web interface
CRC errors in your parser Compact: CRC32 over the whole frame; MSGPACK: CRC32 over the MSGPACK buffer only (MSGPACK Format)
Fewer points than expected Filters and scan range reduce or zero out points
No ROS 2 topics visible on a device with Native ROS 2 support Cyclone DDS configured with the host interface IP; matching domain ID (Native ROS 2)
WebSocket closes with error 1002 on a device with WebSocket support Check the negotiated extensions and firmware; use a client compatible with the negotiated mode (WebSocket Events)
Reduced range or warnings Check contamination indication and clean the optics cover

If the problem persists, create a diagnostic file and contact SICK.

Was this page helpful?